FDA Adverse Event Malfunction Summary report: N

REPAIR CORE IMPACTION DRILL

MDR report key: 1873419 · Received September 17, 2010

Report

Report Number
1811755-2010-01188
Event Type
Malfunction
Date Received
September 17, 2010
Date of Event
August 13, 2010
Report Date
August 23, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS CONFIRMED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH A DAMAGED NOSE CONE, SPINDLE HOUSING, DRIVESHAFT ASSEMBLY AND BEARINGS, WHICH WERE EACH REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THERE WAS NO REPORTED PT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REPAIR CORE IMPACTION DRILL BONE CUTTING INSTRUMENT AND ACCESSORIES DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK