FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 1873416
·
Received September 17, 2010
Report
- Report Number
- 1811755-2010-01209
- Event Type
- Malfunction
- Date Received
- September 17, 2010
- Date of Event
- August 24, 2010
- Report Date
- August 24, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE HAS BEEN RECEIVED AT THE MANUFACTURER FOR INVESTIGATION. AN EVALUATION WAS CONDUCTED AND THE COMPLAINT WAS CONFIRMED. ACCORDING TO THE INVESTIGATION DETAILS, THE BEARINGS WERE BROKEN IN THE SPINDLE HOUSING. THE BEARINGS, MOTOR, DRIVESHAFT, AND SPINDLE HOUSING WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING AN ORAL SURGERY. THERE WAS NO REPORTED PT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | BONE CUTTING INSTRUMENT AND ACCESSORIES | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |