FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 1873266 · Received October 18, 2010

Report

Report Number
3005075853-2010-05920
Event Type
Malfunction
Date Received
October 18, 2010
Date of Event
September 22, 2010
Report Date
September 24, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED WITH JAWS IN THE CLOSED POSITION. THE TRIGGER WAS MANUALLY ASSISTED IN ORDER TO OPEN THE JAWS; ONE PEAR SHAPED CLIP WAS RELEASED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE REMAINING CLIPS WERE CONFORMING ACCORDING TO OUR MANUFACTURING SPECIFICATIONS. IT SHOULD BE NOTED THAT AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE OPENING ISSUES. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS NOT COMPLETELY VISIBLE. THIS FINDING IS NOT RELATED TO THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A RADICAL PROSTATECTOMY PROCEDURE, THE JAW WOULD NO LONGER OPEN. THE JAW DID NOT CLAMP THE TISSUE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK