AUTOMATED PD SET W/CASSETTE4 PRONG
Report
- Report Number
- 1423500-2010-04564
- Event Type
- Malfunction
- Date Received
- October 18, 2010
- Date of Event
- September 24, 2010
- Report Date
- September 24, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE WAS DISCARDED. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
(B)(4). AN ENGINEERING QUALITY REVIEW WAS COMPLETED FOR THIS REPORT OF A SYSTEM ERROR (SE) 2240 ALARM. THE REPORT WAS NOT CONFIRMED DUE TO LACK OF SAMPLE. THE LOT NUMBER WAS UNKNOWN; THEREFORE, A BATCH REVIEW WAS NOT PERFORMED. BASED ON THE INFORMATION OBTAINED FROM BAXTER?S INVESTIGATION, THE ROOT CAUSE OF THE SE 2240 WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE BECAUSE THE BAG WAS NOT CONNECTED TIGHT. THE HOME PATIENT STATED THE HEATER BAG CONNECTION WAS LOOSE. THE LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
A HOME PATIENT (HP) CONTACTED GLOBAL TECHNICAL SERVICES REGARDING A SYSTEM ERROR (SE) 2240 ALARM THAT OCCURRED ON THE HOMECHOICE (HC) UNIT DURING INITIAL DRAIN. THE HP STATED THE HEATER BAG CONNECTION WAS LOOSE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED SE 2240 INDICATES A LARGE AMOUNT OF AIR HAD ENTERED CASSETTE. THE TSR ADVISED THE HP TO CYCLE POWER TO CLEAR THE ALARM AND ASSISTED THE HP TO END THERAPY TO START OVER WITH NEW SUPPLIES. THE TSR REVIEWED PROPER PROCEDURES PER THE USER MANUAL WITH THE HP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.
THE CUSTOMER REPORTED THAT THE 2800 SYSTEM LEG EXTENSION WAS STUCK AND THERE WAS A BURNING SMELL COMING FROM THE SYSTEM. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOMATED PD SET W/CASSETTE4 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR |