CELL DYN EMERALD
Report
- Report Number
- 2919069-2010-00478
- Event Type
- Malfunction
- Date Received
- October 18, 2010
- Date of Event
- October 11, 2010
- Report Date
- October 11, 2010
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION
- Product Code
- GKZ
- PMA / PMN Number
- K081495
- Removal / Correction Number
- 2919069-10/15/10-007-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). CELL-DYN EMERALD CPU BOARDS, PART NUMBERS: 8701686901 AND 8701686902. THE CAUSE OF THE CELL-DYN EMERALD INITIALIZATION (POWERING UP) ISSUE WAS DUE TO A PROBLEM WITH HOW THE FLASH MEMORY IS INITIALIZED ON THE CELL-DYN EMERALD CPU BOARD. ABBOTT ISSUED A PRODUCT CORRECTION LETTER TO ALL CELL-DYN EMERALD CUSTOMERS WHO RECEIVED THE AFFECTED ANALYZERS, INSTRUCTING CUSTOMERS TO INSTALL A PRINTER DRIVER WHICH WILL ELIMINATE THE POSSIBILITY FOR THE ERROR TO OCCUR.
THE CELL-DYN EMERALD IS AN AUTOMATED HEMATOLOGY ANALYZER INTENDED FOR IN VITRO DIAGNOSTIC USE IN THE CLINICAL LABORATORY. AFTER INSTALLATION OF THE ANALYZER, SOME CELL-DYN INSTRUMENTS WOULD NOT INITIALIZE AND POWER UP. THE ERROR MESSAGE GENERATED IS "NO MEMORY AVAILABLE." UPON FURTHER EVALUATION BY THE SUPPLIER, IT WAS DETERMINED THE CELL-DYN EMERALD CENTRAL PROCESSING UNIT (CPU) BOARD APPEARED TO NOT HAVE ENOUGH MEMORY TO INITIALIZE AND POWER UP THE INSTRUMENT DUE TO A PROBLEM WITH HOW THE FLASH MEMORY IS INITIALIZED. IN THIS FAILURE MODE, THE INSTRUMENT IS UNABLE TO GENERATE RESULTS AND HAS THE POTENTIAL TO CAUSE A DELAY IN GENERATING PATIENT RESULTS GREATER THAN ONE HOUR AND UP TO 24 HOURS. FOR CERTAIN ASSAYS, THIS COULD IMPACT PATIENT MANAGEMENT. A PRODUCT CORRECTION HAS BEEN ISSUED AND REPORTED UNDER 21CFR806 FOR THE CELL-DYN EMERALD TO THE FDA (B)(4) ON (B)(4) 2010. ABBOTT HAS NOT RECEIVED ANY REPORTS OF ADVERSE EVENTS RELATED TO THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL DYN EMERALD | AUTOMATED HEMATOLOGY ANALYZER | GKZ | ABBOTT DIAGNOSTICS DIVISION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CELL-DYN EMERALD ANALYZER CPU BOARD |