FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 18729249 · Received February 19, 2024

Report

Report Number
1221359-2024-00194
Event Type
Malfunction
Date Received
February 19, 2024
Date of Event
January 25, 2024
Report Date
February 19, 2024
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

FDA UDI ¿ (B)(6) TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 226152 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT 226152 AND TEST BASE PART NUMBER 195-430H/ LOT 222445. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 226152 SHOWED THAT THE COMPLAINT RATE IS(B)(4) ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE, HOWEVER, IT COULD HAVE POSSIBLY BEEN RELATED TO THE SPECIFIC PATIENT SAMPLE. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED TWO FALSE POSITIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6)2024 AND (B)(6)2024. THIS MANUFACTURER'S REPORT ADDRESSES TEST ONE (1) OF TWO (2). CONSUMER PERFORMED THE FIRST BINAXNOW TEST ON 25JAN2024 AND THE SECOND BINAXNOW TEST ON (B)(6)2024, BOTH GENERATED POSITIVE RESULTS. PCR CONFIRMATION TESTING (PLATFORM UNKNOWN) WAS PERFORMED AT A CLINIC ON 27JAN2024 AND GENERATED A NEGATIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
495174 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 226152 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 NA Male