FDA Adverse Event
Malfunction
Summary report: N
ALLURA XPER FD20
MDR report key: 1872924
·
Received October 5, 2010
Report
- Report Number
- 3003768277-2010-00239
- Event Type
- Malfunction
- Date Received
- October 5, 2010
- Date of Event
- August 18, 2010
- Report Date
- August 18, 2010
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- IZI
- PMA / PMN Number
- K031333
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(METHOD, RESULTS, CONCLUSIONS) - THE INVESTIGATION IS STILL ONGOING ON THIS EVENT. WHEN THE INVESTIGATION IS COMPLETED, A F/U REPORT WILL BE SENT TO THE FDA.
Description of Event or Problem · 1
THIS X-RAY SYSTEM LOST FLUORO AND CINE IMAGE WHILE THE PT WAS ON THE TABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALLURA XPER FD20 | IZI (SYSTEM, X-RAY, ANGIOGRAPHIC) | IZI | PHILIPS MEDICAL SYSTEMS | 722006 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |