FDA Adverse Event Malfunction Summary report: N

AFFINITY 3

MDR report key: 1872715 · Received October 8, 2010

Report

Report Number
1824206-2010-10385
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
September 15, 2010
Report Date
September 15, 2010
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT ONE STRAP WAS MISSING AND THE OTHER LOOSE ON THE LOWER LIFT ARM. THIS MAY HAVE CAUSE THE FRAME TO TWIST SLIGHTLY, HITTING LIFT ARM AND AFFECTING TREND. THE TECHNICIAN REPLACED THE RETAINING STRAP WITH A NEW BUSHING AND SCREWS TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BED IS NOT GOING INTO TRENDELENBURG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 3 BIRTHING BED HDD HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1