FDA Adverse Event
Malfunction
Summary report: N
AFFINITY 3
MDR report key: 1872715
·
Received October 8, 2010
Report
- Report Number
- 1824206-2010-10385
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- September 15, 2010
- Report Date
- September 15, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THAT ONE STRAP WAS MISSING AND THE OTHER LOOSE ON THE LOWER LIFT ARM. THIS MAY HAVE CAUSE THE FRAME TO TWIST SLIGHTLY, HITTING LIFT ARM AND AFFECTING TREND. THE TECHNICIAN REPLACED THE RETAINING STRAP WITH A NEW BUSHING AND SCREWS TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE BED IS NOT GOING INTO TRENDELENBURG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY 3 | BIRTHING BED | HDD | HILL-ROM, INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |