FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL

MDR report key: 1872698 · Received October 18, 2010

Report

Report Number
6000001-2010-04141
Event Type
Malfunction
Date Received
October 18, 2010
Date of Event
August 11, 2010
Report Date
August 12, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 814:04 WAS CONFIRMED BUT NOT DUPLICATED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS NOT IDENTIFIED. NO FURTHER TESTING HAS BEEN COMPLETED AT THIS TIME AND NO REPAIRS HAVE BEEN PERFORMED AS THE REFERENCED DEVICE HAS BEEN REMOVED FROM SERVICE. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 814:04. ACCORDING TO THE FACILITY, IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. UPON REVIEWING THE PUMP'S EVENT HISTORY, BAXTER QUALITY ENGINEERING DISCOVERED THIS EVENT INTERRUPTED DELIVERY ON CHANNEL A. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90.