FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1872572 · Received October 8, 2010

Report

Report Number
2531779-2010-01534
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
September 8, 2010
Report Date
September 8, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED THAT THE KEYPAD WAS PEELING AT THE "UP" BUTTON. THE KEYPAD BUTTONS WERE REPORTEDLY RESPONSIVE AND SPRINGING BACK NORMALLY. EVIDENCE OF ADHESIVE AND MOISTURE CORROSION WAS OBSERVED UNDER BUTTON CONTACTS. THE KEYPAD WAS OBSERVED TO BE PEELING DURING INVESTIGATION. CONTAMINATION WAS OBSERVED UNDER KEYPAD.

Description of Event or Problem · 1

THE PT ALLEGED THAT A PUMP WAS LESS RESPONSIVE TO PRESSING OF THE UP ARROW BUTTON. THE PT DENIED ANY BG EXCURSION ASSOCIATED WITH PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. IR 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 16 YR