ANIMAS INSULIN INFUSION PUMP
MDR report key: 1872569
·
Received October 8, 2010
Report
- Report Number
- 2531779-2010-01537
- Event Type
- Malfunction
- Date Received
- October 8, 2010
- Date of Event
- September 10, 2010
- Report Date
- September 10, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO REPORTED PT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED A MISALIGNED DISPLAY SCREEN AND DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALED LOSS OF PRIME WARNINGS ASSOCIATED WITH ZERO FORCE AND "NO CARTRIDGE DETECTED" WARNINGS. DURING EVAL A LOAD CARTRIDGE STEP WAS PERFORMED AND THE PUMP DID NOT DETECT THE CARTRIDGE, DISPENSED FLUID FROM THE CARTRIDGE, AND EMITTED A "NO CARTRIDGE DETECTED" WARNING.
Description of Event or Problem · 1
THERE WAS NO REPORTED PT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE DISTRIBUTOR REPORTED THAT THE PUMP DISPENSED INSULIN FROM THE CARTRIDGE DURING AN "EZPRIME" OPERATION.