FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1872569 · Received October 8, 2010

Report

Report Number
2531779-2010-01537
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
September 10, 2010
Report Date
September 10, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO REPORTED PT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED A MISALIGNED DISPLAY SCREEN AND DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALED LOSS OF PRIME WARNINGS ASSOCIATED WITH ZERO FORCE AND "NO CARTRIDGE DETECTED" WARNINGS. DURING EVAL A LOAD CARTRIDGE STEP WAS PERFORMED AND THE PUMP DID NOT DETECT THE CARTRIDGE, DISPENSED FLUID FROM THE CARTRIDGE, AND EMITTED A "NO CARTRIDGE DETECTED" WARNING.

Description of Event or Problem · 1

THERE WAS NO REPORTED PT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE DISTRIBUTOR REPORTED THAT THE PUMP DISPENSED INSULIN FROM THE CARTRIDGE DURING AN "EZPRIME" OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1