FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 1872349 · Received October 8, 2010

Report

Report Number
9615739-2010-00234
Event Type
Malfunction
Date Received
October 8, 2010
Date of Event
September 15, 2010
Report Date
September 15, 2010
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN TRACED THE ISSUE TO THE SIDECOMM BOARD. THE TECHNICIAN REPLACED THE SIDECOMM BOARD TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION REC'D INDICATES THE BED EXIT WILL SET AND ALARM LOCALLY BUT WILL NOT ALARM AT THE NURSE STATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC POWERED HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1