FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 1872296
·
Received October 12, 2010
Report
- Report Number
- 1720753-2010-03514
- Event Type
- Malfunction
- Date Received
- October 12, 2010
- Date of Event
- September 23, 2010
- Report Date
- October 12, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REPRESENTATIVE EVALUATED THE SYSTEM AND COULD NOT REPRODUCE THE REPORTED PROBLEM. COLLIMATOR WAS CLOSE TO BEING OUT OF SPECIFICATION, SO A COLLIMATOR CALIBRATION WAS PERFORMED. SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
CUSTOMER REPORTED THE SYSTEM IS DISPLAYING A COLLIMATOR IRIS TOO LARGE ERROR. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |