FDA Adverse Event Malfunction Summary report: N

BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE

MDR report key: 18722646 · Received February 16, 2024

Report

Report Number
9617032-2024-00204
Event Type
Malfunction
Date Received
February 16, 2024
Date of Event
February 1, 2024
Report Date
May 16, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903643146
PMA / PMN Number
K022426
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E1. INITIAL REPORTER PHONE #: (B)(6). E1. INITIAL REPORTER FACILITY NAME: (B)(6). H.6. INVESTIGATION SUMMARY: MATERIAL #: 364314 LOT/BATCH #: 3122762 BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR FOREIGN MATTER WAS OBSERVED. ADDITIONALLY, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF FOREIGN MATTER WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE FOREIGN MATTER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. H3 OTHER TEXT : SEE H.10

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO A CORRECTION: B3: DATE OF EVENT: 01-FEB-2024.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USING BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE, THE CANNULA WAS STAINED. NO PATIENT IMPACT REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USING BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE, THE CANNULA WAS STAINED. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
509094 BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 3122762 50382903643146

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown