D-VAPOR
Report
- Report Number
- 9611500-2010-00035
- Event Type
- Other
- Date Received
- August 27, 2010
- Date of Event
- July 26, 2010
- Report Date
- August 27, 2010
- Manufacturer
- DRAEGER MEDICAL AG & CO. KG
- Product Code
- CAD
- PMA / PMN Number
- K042276
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- UNKNOWN
Narratives
NOTE: THIS REPORT WAS FILED IN RESPONSE TO VOLUNTARY UFR, ESPECIALLY WITH RESPECT TO THE UF STATEMENT/GUESS THAT THE AFFECTED POWER CORD OF THE DRAEGER D-VAPOR IS SUBJECT OF AN ONGOING PRODUCT RECALL OF POWER CORDS (MADE BY ELECTRI-CORD). THE AFFECTED DEVICE (INCL. POWER CORD) WAS REQUESTED FOR INVESTIGATION AT MFR SITE AND WAS RECEIVED (B)(4) 2010. IT CAN BE STATED THAT THE POWER CORD IN QUESTION IS NOT SUBJECT TO A PRODUCT RECALL OF POWER CORDS (MADE BY ELECTRI-CORD). THE ELECTRICAL PLUG IDENTIFIED IN THIS CASE WAS MFG BY FELLER LLC. THE POWER CORD WAS SUBMITTED TO FELLER LLC FOR FURTHER INVESTIGATION; RESULTS ARE PENDING. DRAEGER MUST EMPHASIZE THAT THIS IS THE FIRST REPORTED CASE ON A BURNT POWER PLUG OF THIS TYPE. THE INVESTIGATION IS ONGOING; RESULTS WILL BE REPORTED IN A FOLLOW-UP REPORT.
MFR RECEIVED A VOLUNTARY MEDWATCH REPORT FROM A USER FACILITY ((B)(6)) REGARDING A DRAEGER DESFLURANE VAPORIZER. BY THIS VOLUNTARY UFR IT IS REPORTED THAT THE ANESTHESIOLOGIST NOTICED THAT THE VAPORIZER DID NOT POWER ON. IT WAS FOUND THAT THE NEUTRAL PIN OF THE ELECTRICAL POWER PLUG WAS BURNED OFF AND WAS STUCK IN THE POWER RECEPTACLE. IT WAS FURTHER STATED IN THIS VOLUNTARY UFR THAT THE DEVICE IN QUESTION HAD AN ELECTRI-CORD POWER CORD, WHICH IS SUBJECT OF OTHER MFR'S RECALL, BUT NOT BY DRAEGER AS OF YET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | D-VAPOR | VAPORIZER, ANESTHETIC AGENT | CAD | DRAEGER MEDICAL AG & CO. KG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |