FDA Adverse Event Malfunction Summary report: N

ETS FLEX

MDR report key: 1872237 · Received October 18, 2010

Report

Report Number
3005075853-2010-05898
Event Type
Malfunction
Date Received
October 18, 2010
Date of Event
August 25, 2010
Report Date
September 23, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE CARTRIDGE WAS LOADED INTO THE DEVICE WITHOUT ANY DIFFICULTIES AND DID NOT FELL OUT DURING FUNCTIONAL TESTING. EVENT COULD NOT BE CONFIRMED AS NO CARTRIDGE WAS RECEIVED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE'S CARTRIDGE FELL OFF OF THE STAPLER AFTER IT WAS LOADED OUTSIDE OF THE PATIENT. THE CARTRIDGE DID NOT FALL INTO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. (B)(4)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETS FLEX STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK NI

Patients

Seq Age Sex Outcome Treatment
1