DECANAV ELECTROPHYSIOLOGY CATHETER

MDR report key: 18721959 · Received February 16, 2024

Report

Report Number
2029046-2024-00518
Event Type
Malfunction
Date Received
February 16, 2024
Date of Event
January 22, 2024
Report Date
March 21, 2024
Manufacturer
BIOSENSE WEBSTER INC
Product Code
DRF
UDI-DI
10846835008807
PMA / PMN Number
K080425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

IT WAS REPORTED THAT A DECANAV ELECTROPHYSIOLOGY CATHETER WAS ORDERED FROM BIOSENSE WEBSTER INC. BY STERILMED INC. TO USE AS A BASELINE REFERENCE IN A TEST BEING CONDUCTED. THE CATHETER WAS REVIEWED UNDER A MICROSCOPE AND THE EDGE OF ONE OF THE ELECTRODES WAS EXPOSED, AS WELL AS THE WIRE UNDER IT. IT WAS BELIEVED THAT THERE WAS NOT ENOUGH GLUE TO PROPERLY COVER IT LEADING TO A SHARP EDGE. THERE WAS NO PATIENT INVOLVEMENT. DEVICE EVALUATION DETAILS: THE DEVICE WAS RETURNED TO BIOSENSE WEBSTER (BWI) FOR EVALUATION. A VISUAL AND DIMENSIONAL INSPECTION AND DEFLECTION TEST OF THE RETURNED DEVICE WAS PERFORMED IN ACCORDANCE WITH BWI PROCEDURES. VISUAL ANALYSIS REVEALED NO DAMAGE OR ANOMALIES ON THE DEVICE. THE DISTAL TIP AND THE ELECTRODES WERE REVIEWED IN DETAIL AND NO ANOMALIES WERE FOUND. A DEFLECTION TEST WAS PERFORMED, AND THE CURVE WAS DEFLECTING WITHIN SPECIFICATIONS. NO DEFLECTION ISSUES WERE OBSERVED. A DIMENSIONAL INSPECTION WAS PERFORMED, AND THE DEVICE PASSED WITHIN SPECIFICATIONS. NO ISSUE WAS IDENTIFIED DURING THE ANALYSIS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE NUMBER LOT 31199385M, AND NO INTERNAL ACTION RELATED TO THE COMPLAINT WAS FOUND DURING THE REVIEW. ADDITIONALLY, THE MANUFACTURE AND EXPIRATION DATES HAVE BEEN PROVIDED. THEREFORE, FIELDS D4. EXPIRATION DATE AND H4. DEVICE MANUFACTURE DATE HAVE BEEN POPULATED. THE ISSUE REPORTED BY THE CUSTOMER WAS NOT CONFIRMED. NO MANUFACTURING ISSUE WERE IDENTIFIED DURING THE ANALYSIS. OTHER ISSUES OR CIRCUMSTANCES MAY HAVE OCCURRED DURING THE USAGE OF THE DEVICE THAT COMPROMISED ITS PERFORMANCE. AS PART OF BIOSENSE WEBSTER¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A DECANAV ELECTROPHYSIOLOGY CATHETER WAS ORDERED FROM BIOSENSE WEBSTER INC. BY STERILMED INC. TO USE AS A BASELINE REFERENCE IN A TEST BEING CONDUCTED. THE CATHETER WAS REVIEWED UNDER A MICROSCOPE AND THE EDGE OF ONE OF THE ELECTRODES WAS EXPOSED, AS WELL AS THE WIRE UNDER IT. IT WAS BELIEVED THAT THERE WAS NOT ENOUGH GLUE TO PROPERLY COVER IT LEADING TO A SHARP EDGE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
416152 DECANAV ELECTROPHYSIOLOGY CATHETER CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING DRF BIOSENSE WEBSTER INC 31199385M 10846835008807

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown