FDA Adverse Event
Malfunction
Summary report: N
INFINITI VISION SYSTEM OZIL
MDR report key: 1872165
·
Received October 14, 2010
Report
- Report Number
- 2028159-2010-01937
- Event Type
- Malfunction
- Date Received
- October 14, 2010
- Date of Event
- September 14, 2010
- Report Date
- September 14, 2010
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K082845
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY SERVICE REP EXAMINED THE SYSTEM AND WAS UNABLE TO DUPLICATE THE PROBLEM REPORTED. PREVENTIVE MAINTENANCE WAS PERFORMED. THE SOFTWARE WAS UPDATED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE I/A FUNCTION WAS VERY SLOW AND WEAK. THERE WERE NO SYSTEM MESSAGES DISPLAYED. ADD'L INFO WAS REQUESTED ON THE EVENT AND PT STATUS WITH NO RESPONSE TO DATE. THE PT IMPACT IS UNK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITI VISION SYSTEM OZIL | PHACOFRAGMENTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CENTER | INFINITI OZIL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |