FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1872145
·
Received October 11, 2010
Report
- Report Number
- 2531779-2010-01551
- Event Type
- Malfunction
- Date Received
- October 11, 2010
- Date of Event
- September 12, 2010
- Report Date
- September 12, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THERE WAS NO DAMAGE OBSERVED TO THE KEYPAD, THE PUMP INTERMITTENTLY RESPONDED TO BUTTON PRESSES, AND THERE WAS ADHESIVE/CONTAMINATION FOUND UNDER THE BUTTON CONTACTS.
Description of Event or Problem · 1
THE PT'S MOTHER CLAIMED THAT THE OK BUTTON STARTED STICKING 2 WEEKS AGO AND THAT IT REQUIRED MULTIPLE PRESSES FOR THE PUMP TO RESPOND. THE REPORTER DENIES STRUCTURAL DAMAGE TO THE PUMP BUT INDICATED THAT THE PAINT HAS WORN OFF THE BUTTONS. THE PUMP REPORTEDLY STILL DELIVERS INSULIN AS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | IR 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |