FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1872144 · Received October 11, 2010

Report

Report Number
2531779-2010-01553
Event Type
Malfunction
Date Received
October 11, 2010
Date of Event
September 12, 2010
Report Date
September 12, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: ALL BUTTONS CLICK AND SPRING BACK NORMALLY, THERE WAS NO DAMAGE OBSERVED TO THE KEYPAD, THE PUMP INTERMITTENTLY RESPONDED TO UP ARROW BUTTON PRESSES, AND THERE WAS ADHESIVE/CONTAMINATION FOUND UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PT'S MOTHER CLAIMED THAT THE UP ARROW BUTTON STARTED INTERMITTENTLY STICKING 1-2 WEEKS AGO, REQUIRES MULTIPLE PRESSES FOR THE PUMP TO RESPOND. THE REPORTER INDICATED THAT THE BUTTON FEELS FLAT WHEREAS THE OTHER BUTTONS FEEL NORMAL. THE REPORTER DENIES WATER EXPOSURE OR CLEANING PRODUCTS. THE PUMP REPORTEDLY STILL DELIVERS INSULIN AS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 17 YR