FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1872144
·
Received October 11, 2010
Report
- Report Number
- 2531779-2010-01553
- Event Type
- Malfunction
- Date Received
- October 11, 2010
- Date of Event
- September 12, 2010
- Report Date
- September 12, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: ALL BUTTONS CLICK AND SPRING BACK NORMALLY, THERE WAS NO DAMAGE OBSERVED TO THE KEYPAD, THE PUMP INTERMITTENTLY RESPONDED TO UP ARROW BUTTON PRESSES, AND THERE WAS ADHESIVE/CONTAMINATION FOUND UNDER THE BUTTON CONTACTS.
Description of Event or Problem · 1
THE PT'S MOTHER CLAIMED THAT THE UP ARROW BUTTON STARTED INTERMITTENTLY STICKING 1-2 WEEKS AGO, REQUIRES MULTIPLE PRESSES FOR THE PUMP TO RESPOND. THE REPORTER INDICATED THAT THE BUTTON FEELS FLAT WHEREAS THE OTHER BUTTONS FEEL NORMAL. THE REPORTER DENIES WATER EXPOSURE OR CLEANING PRODUCTS. THE PUMP REPORTEDLY STILL DELIVERS INSULIN AS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |