FDA Adverse Event Malfunction Summary report: N

BIPOLAR LEAD MODEL 300

MDR report key: 1872140 · Received October 14, 2010

Report

Report Number
1644487-2010-02297
Event Type
Malfunction
Date Received
October 14, 2010
Date of Event
September 14, 2010
Report Date
September 15, 2010
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

DEVICE MALFUNCTION SUSPECTED BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

A NEUROLOGY OFFICE REPORTED TO OUR COUNTRY REPRESENTATIVE THAT THEY HAD A VNS PATIENT WHO WAS HAVING LONGER AND INCREASED SEIZURES NOT ABOVE THEIR PREVNS BASELINE RATE. A SYSTEM DIAGNOSTIC TEST WAS PERFORMED AND SHOWED THE PATIENT TO HAVE HIGH LEAD IMPEDANCE. THE PATIENT'S TREATING PHYSICIAN'S OFFICE BELIEVES THE CAUSE OF THE PATIENT'S SEIZURES IS THEIR HIGH LEAD IMPEDANCE. THE PATIENT IS GOING TO BE SCHEDULED FOR REVISION SURGERY. NO DATE HAS BEEN SET AT THIS TIME. THE PATIENT'S VNS REMAINS PROGRAMMED ON. GOOD FAITH ATTEMPTS ARE UNDERWAY FOR FURTHER INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BIPOLAR LEAD MODEL 300 LYJ CYBERONICS, INC. 300-20 5760

Patients

Seq Age Sex Outcome Treatment
1 15 YR