FDA Adverse Event Injury Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1872048 · Received October 8, 2010

Report

Report Number
2032227-2010-82876
Event Type
Injury
Date Received
October 8, 2010
Date of Event
September 20, 2010
Report Date
September 20, 2010
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 455 MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME, DISPLACEMENT AND SELF TEST. REVIEWED THE ALARM HISTORY, AND FOUND MULTIPLE MOTOR ERROR ALARMS. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization