FDA Adverse Event Malfunction Summary report: N

COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH

MDR report key: 1872020 · Received October 18, 2010

Report

Report Number
6000001-2010-04099
Event Type
Malfunction
Date Received
October 18, 2010
Date of Event
June 23, 2010
Report Date
September 23, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A BATTERY DEPLETED SET ALARM WAS CONFIRMED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. THIS DEVICE IS A COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.06.00. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER (B)(4) PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A DEPLETED BATTERY ALARM, WHICH INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS DEVICE IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1