FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 1872018
·
Received October 14, 2010
Report
- Report Number
- 1720753-2010-03566
- Event Type
- Malfunction
- Date Received
- October 14, 2010
- Date of Event
- September 30, 2010
- Report Date
- October 14, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE BOARDS AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9800 SYSTEM LOCKED UP WITH A BLANK SCREEN AND HIGH MA ERROR DURING A CASE. THE 9800 SYSTEM HAD TO BE REBOOTED AND THE PROCEDURE WAS COMPLETED WITHOUT INCIDENT. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |