FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 CXE VOLUMETRICINFUSION PUMP

MDR report key: 1872012 · Received October 18, 2010

Report

Report Number
6000001-2010-04094
Event Type
Malfunction
Date Received
October 18, 2010
Date of Event
April 12, 2010
Report Date
September 23, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: UPON REVIEW OF THIS EVENT, IT WAS FOUND THAT DUPLICATE REPORTS WERE OPENED FOR A BATTERY DEPLETED ALARM THAT OCCURRED IN ALL THREE CHANNELS (A, B AND C). THE BATTERY DEPLETED ALARM IS A SYSTEM FAILURE ON THE PUMP AND DOES NOT WARRANT SEPARATE REPORTS PER CHANNEL. BASED ON THIS REVIEW, THE INITIAL MEDWATCH IS A DUPLICATE OF MANUFACTURERS REPORT NUMBER 6000001-2010-04093.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DEPLETED BATTERY ALARM ON CHANNEL C WAS FOUND AND CONFIRMED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY DEPLETED AND DAMAGED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER (B)(4) PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A DEPLETED BATTERY ALARM, WHICH INTERRUPTED DELIVERY ON CHANNEL C. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.92.