IMPLANTABLE CHF PULSE GENERATOR
Report
- Report Number
- 2124215-2010-17499
- Date Received
- October 16, 2010
- Report Date
- September 29, 2010
- Product Code
- NIK
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
AS OF TODAY, THIS PRODUCT HAS NOT BEEN RETURNED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.
UPON RECEIPT AT THE POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION REVEALED THAT THE HEADER WAS LOOSE; HOWEVER STILL ATTACHED TO THE CAN. AN X-RAY WAS PERFORMED WHICH INDICATED THAT THE FEED THRU WIRES WERE BROKEN. THE REPORTED CLINICAL OBSERVATIONS WERE A DIRECT RESULT OF THE BROKEN FEED THRU WIRES CAUSED BY A LOOSE HEADER. DUE TO THE BROKEN FEED THRU WIRES, NO THERAPY WAS AVAILABLE FOR THIS PRODUCT WAS THE TIME OF EXPLANT. IMPROVEMENTS HAVE BEEN MADE TO THE MANUFACTURING PROCESS IN ORDER TO STRENGTHEN THE BOND BETWEEN THE HEADER AND THE DEVICE CASE.
ANALYSIS IS CURRENTLY ON-GOING. UPON COMPLETION OF ANALYSIS THIS REPORT WILL BE UPDATED.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THE PATIENT WITH THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) HAD RECEIVED A SHOCK. IT WAS FURTHER REPORTED THAT AT THEIR LAST FOLLOW-UP APPOINTMENT THE PHYSICIAN BELIEVED THAT THERE WAS A LOOSE CONNECTION ISSUE OR A RIGHT VENTRICULAR (RV) LEAD FRACTURE AS THERE WAS NOISE ON THE RATE/SENSE PORTION DURING ISOMETRICS IN ADDITION TO OUT OF RANGE IMPEDANCE MEASUREMENTS ON BOTH THE PACING AND SHOCKING PORTIONS. THE PATIENT UNDERWENT A REVISION PROCEDURE AND THE ROOT CAUSE WAS UNABLE TO BE DETERMINED DURING TROUBLESHOOTING. A NEW RV LEAD WAS FIRST PLACED AND IMPEDANCES REMAINED OUT OF RANGE. THE CRT-D WAS THEN ALSO REPLACED AND NO FURTHER COMPLICATIONS WERE REPORTED. NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED. THIS PRODUCT WAS TO BE RETURNED FOR LABORATORY ANALYSIS.
--
ADDITIONAL INFORMATION INDICATES THAT THIS PRODUCT WAS RETURNED FOR LABORATORY ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMPLANTABLE CHF PULSE GENERATOR | NIK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0185| 4555| N119| 4470 |