FDA Adverse Event Malfunction Summary report: N

TELIGEN

MDR report key: 1871921 · Received October 16, 2010

Report

Report Number
2124215-2010-17449
Event Type
Malfunction
Date Received
October 16, 2010
Date of Event
July 14, 2010
Report Date
January 6, 2017
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
PMA / PMN Number
P960040
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE REMAINS IN SERVICE.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT DURING IMPLANT THIS LEAD EXHIBITED NOISE AND LOW SHOCK IMPEDANCES. THE LEAD WAS REMOVED AND RESEATED IN THE HEADER WHICH RESOLVED THE ISSUE.

Description of Event or Problem · 1

APPROXIMATELY SIX YEARS LATER THIS DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWP GUIDANT CRM CLONMEL IRELAND E110

Patients

Seq Age Sex Outcome Treatment
1 72 YR