TELIGEN
Report
- Report Number
- 2124215-2010-17449
- Event Type
- Malfunction
- Date Received
- October 16, 2010
- Date of Event
- July 14, 2010
- Report Date
- January 6, 2017
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- PMA / PMN Number
- P960040
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE DEVICE REMAINS IN SERVICE.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT DURING IMPLANT THIS LEAD EXHIBITED NOISE AND LOW SHOCK IMPEDANCES. THE LEAD WAS REMOVED AND RESEATED IN THE HEADER WHICH RESOLVED THE ISSUE.
APPROXIMATELY SIX YEARS LATER THIS DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS.