HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2023-40794
- Event Type
- Malfunction
- Date Received
- February 16, 2024
- Date of Event
- February 5, 2023
- Report Date
- February 15, 2024
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 00730002856789
- PMA / PMN Number
- K181216
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT HAS BEEN REOPENED AND REVIEWED ACCORDING TO FDA FORM 483 INSPECTIONAL OBSERVATION EMS #2, EOBS2 FROM THE FDA INSPECTION CONDUCTED BETWEEN JULY 17 TO JULY 21, 2022 AT THE EMS AND BONADUZ SITES. A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG ALMOST 2 YEARS AGO, NO ATTEMPTS WILL BE PERFORMED TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS PREPARED FOR TREATMENT. THE ROOT CAUSE COULD NOT BE FULLY DETERMINED AS WHEN THE TOUCHSCREEN IS NOT FULLY RESPONSIBLE THEN THE ROOT CAUSE COULD BE IN THE WHOLE CHAIN. DEFECTIVE TOUCHSCREEN-> DEFECTIVE FRONT PANEL BOARD. FLAT CABLE ISSUE BETWEEN FRONT PANEL BOARD AND CONTROL BOARD AND THE ESM (SPI CHANNEL) IN CONSEQUENCE THE CORRECTION TO REPLACE MSP 161290 DISPLAY FRONT COMPLETE CORRECTED THE ISSUE. THERE WAS NO PATIENT OR USER HARM.
DEAR MILAD, A CUSTOMER HAD A PROBLEM WITH A C1: THE DISPLAY FRONT P.N. MSP161290 S.N. (B)(6) IS DEFECTIVE: TOUCH SCREEN DOESN'T REACT. WE REPLACED THE PART AND DID ALL THE TESTS. ICHILOV SC2300150 B.R. ILAN OFMAN BEPEX LTD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1932821 | HAMILTON MEDICAL AG | HAMILTON-C1 | CBK | HAMILTON MEDICAL AG | HAMILTON-C1 | 00730002856789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |