DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2010-00442
- Event Type
- Other
- Date Received
- October 15, 2010
- Date of Event
- September 16, 2010
- Report Date
- September 16, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081137
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING (FSE) CONCLUDED THAT THE ISSUE EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH THE SURGEON SIDE CONSOLE (SSC) POWER SUPPLY. THE SSC POWER SUPPLY CONVERTS THE AC VOLTAGE FROM THE WALL OUTLET TO DC VOLTAGE FOR USE BY THE SYSTEM. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED POWER SUPPLY. AS OF (B)(4) 2010, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE, THE SITE EXPERIENCED A PATIENT CART RUNNING ON BATTERY MESSAGE DISPLAYED ON THEIR MONITOR. WITH THE ASSISTANCE OF AN ISI TECHNICAL SUPPORT ENGINEER, THE SITE VERIFIED THAT THE ELECTRICAL CABLES WERE PROPERLY PLUGGED INTO THE WALL SOCKET AND THE BREAKER WAS IN THE CORRECT POSITION. THE SITE ALSO REMOVED INSTRUMENTS FROM THE SYSTEM, APPLIED EMERGENCY POWER OFF TO THE SYSTEM AND TURNED OFF THE BREAKER ON PATIENT SIDE CART BEFORE RESTARTING THE SYSTEM. THEY SYSTEM MESSAGE CONTINUED. THE SITE RE-INSTALLED THE INSTRUMENTS BUT COULD NOT ACTIVATE THE INSTRUMENTS FROM SURGEON SIDE CONSOLE. THE MONITOR DISPLAYED THE MESSAGE NO BATTERY BACKUP. AT THIS TIME, THE SURGEON DECIDED TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO COMPLETE THE PLANNED PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000 A6.0P4 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI SI SYSTEM INSTRUMENTS, ACCESSORIES & ESU |