OT PING METER
Report
- Report Number
- 2939301-2010-09120
- Event Type
- Malfunction
- Date Received
- October 15, 2010
- Report Date
- October 9, 2010
- Manufacturer
- LIFESCAN, INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP # 1/SUPPLEMENTAL REPORT TEXT 12/07/2010. THE LAY USER/PATIENT'S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE HAS PASSED TESTING WITH NO FAULTS FOUND. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT; 510 (K) # IS K082590.
IT WAS REPORTED, THE TARGETER IS OLD, AND EVERYTHING PROCEEDED NORMALLY. THE NAIL WAS PASSED CORRECTLY, THE LAG SCREW WAS INSERTED, AND ON INSERTION OF THE DISTAL SCREW, THE DRILL MISSED THE NAIL, THROUGH THE GUIDE AND SCREW WAS PLACED POSTERIOR TO THE NAIL. REMOVED THE SCREW, TRIED TO REDRILL THROUGH THE GUIDE, AND THE DRILL BIT SHATTERED INSIDE THE NAIL. HAD TO REMOVE THE FAILED GAMMA III GUIDE, AND COMPLETED FREEHAND. AFTER SURGERY, REP LOOKED AT IT TO SEE HOW IT LOOKS, THERE WAS JUST SO MUCH PLAY IN IT. THE SHATTERED DRILL BIT WAS REMOVED, BUT PARTICLES WERE LEFT INSIDE.
THE LAY USER/PATIENT CONTACTED LIFESCAN ALLEGING THE METER DOES NOT POWER ON BY PRESSING BUTTONS. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |