OT ULTRALINK METER
Report
- Report Number
- 2939301-2010-09112
- Event Type
- Malfunction
- Date Received
- October 15, 2010
- Report Date
- October 12, 2010
- Manufacturer
- LIFESCAN, INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE METER WAS TESTED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED. HOWEVER, A SECONDARY ISSUE WAS NOTED, THE METER WAS FOUND TO HAVE A LOW/DEAD BATTERY. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT; 510 (K) # IS K073231.
A PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "162 MG/DL" WITH A LIFESCAN METER AND "105 MG/DL" ON A LABORATORY DEVICE, PERFORMED WITHIN 10 MINUTES OF EACH OTHER. BETWEEN THE TESTS THERE WAS NO INTERVENTION THAT WOULD BE EXPECTED TO CHANGE THE BLOOD GLUCOSE. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN'S CRITERIA FOR ACCURACY. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT ULTRALINK METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |