VITALITY 2
Report
- Report Number
- 2124215-2010-17568
- Event Type
- Malfunction
- Date Received
- October 15, 2010
- Date of Event
- August 30, 2010
- Report Date
- September 3, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED WHEN ANALYSIS IS COMPLETE.
UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. ALL LOW VOLTAGE POWER SUPPLY MEASUREMENTS WERE CURRENTLY WITHIN AN APPROPRIATE RANGE; HOWEVER, FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION. THIS DEVICE IS PART OF THE EXPANDED MARCH 2009 POPULATION OF DEVICES DESCRIBED IN THE SHORTENED REPLACEMENT WINDOW ADVISORY. THIS ADVISORY WAS ORIGINALLY COMMUNICATED APRIL 5, 2007. THIS ISSUE IS DISCUSSED IN THE Q3 2010 PRODUCT PERFORMANCE REPORT.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION. INITIAL LABORATORY ANALYSIS FOUND THE DEVICE DID NOT MEET LONGEVITY EXPECTATIONS PER PROGRAMMED VALUES.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 91 YR | T175| 0158| E102 |