FDA Adverse Event Malfunction Summary report: N

ALTRUA

MDR report key: 1871488 · Received October 15, 2010

Report

Report Number
2124215-2010-17148
Event Type
Malfunction
Date Received
October 15, 2010
Date of Event
May 28, 2010
Report Date
August 27, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE WAS NOT SUCCESSFULLY IMPLANTED. THE RIGHT VENTRICULAR LEAD COULD NOT BE INSERTED INTO THE HEADER AS THE SET SCREW WAS SIDEWAYS. ATTEMPTS WERE MADE TO PULL BACK THE SET SCREW, BUT THE SET SCREW WOULD NOT BACK OUT OF THE HEADER. A NEW DEVICE WAS SUCCESSFULLY IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALTRUA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND S603

Patients

Seq Age Sex Outcome Treatment
1