FDA Adverse Event Malfunction Summary report: N

HALF DAY INFUSOR, 5ML/HR 12PK

MDR report key: 1871383 · Received October 15, 2010

Report

Report Number
6000001-2010-04067
Event Type
Malfunction
Date Received
October 15, 2010
Date of Event
September 1, 2010
Report Date
September 22, 2010
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K982102
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: ONE SAMPLE WAS RECEIVED BY BAXTER FOR EVALUATION. A SMALL AMOUNT OF SOLUTION WAS NOTED IN THE BAG THAT CONTAINED THE SAMPLE. THE BLUE WINGED CAP WAS VISUALLY NOTED TO BE ADEQUATELY FASTENED TO THE LUER BODY. DURING THE LEAK TEST, THE SAMPLE WAS ALLOWED TO COMPLETELY PRIME THEN THE BLUE WINGED CAP WAS SECURELY FASTENED TO THE LUER BODY. AFTER 3 DAYS OF MONITORING PERIOD, NO SIGNS OF LEAK WERE OBSERVED ON THE SAMPLE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THIS IS A SINGLE USE DEVICE AND HAS BEEN DISCARDED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR SHOULD ANY FURTHER INFORMATION BECOME AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT ONE (1) HALF DAY INFUSOR DEVICE WAS OBSERVED LEAKING. ACCORDING TO THE REPORT, THE DEVICE IS FILLED WITH DEFERRIOXAMINE (CONCENTRATION 2.5G IN 50ML OF WATER FOR IRRIGATION). THERE IS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HALF DAY INFUSOR, 5ML/HR 12PK PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 10G001

Patients

Seq Age Sex Outcome Treatment
1