FDA Adverse Event Injury Summary report: N

PIEZOSURGERY

MDR report key: 1871368 · Received June 14, 2010

Report

Report Number
3003933619-2010-00001
Event Type
Injury
Date Received
June 14, 2010
Date of Event
March 17, 2010
Report Date
March 22, 2010
Manufacturer
MECTRON S.P.A.
Product Code
ELC
PMA / PMN Number
K052518
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE INSERT OT4 ARRIVED ON 05/13/2010 AT MECTRON FOR INVESTIGATION. THE INSERT WAS BROKEN OFF BUT WE FOUND THE FRAGMENTS. ACCORDING TO THE DENTIST'S COMPLAINT THE INSERT HAD TO BE SURGICALLY REMOVED AT THE IMPLANT SITE FROM A BUCCAL APPROACH. THE PT HAS NOT REPORTED ANY DAMAGE. THE LOT NUMBER OF THE INSERT IS 12/08 CORRESPONDING TO THE MANUFACTURING DATE: DECEMBER 2008. THE RETURNED INSERT WAS VISUALLY INSPECTED AND NO ABNORMALITIES WERE NOTED (NO SCRATCHING AND DEFORMATION WERE FOUND ON THE BROKEN INSERT THAT COULD TRACE THE BREAKAGE BACK TO A MATERIAL CRACK/DEFECT). THE DATE CONTAINED IN THE CASTING CERTIFICATE RELATIVE TO THE MATERIAL BY WHICH THE SUBJECT INSERT LOT WAS MANUFACTURED ARE IN CONFORMITY WITH THE INSERT DESIGN REQUIREMENTS. SEVERAL FACTORS MAY CONTRIBUTE TO THE INSERT BREAKAGE SUCH AS AN EXCESSIVE PRESSURE APPLIED, A CORRECT TECHNIQUE AND LEVERAGE MADE WITH THE INSERT DURING THE OSTEOTOMY. ALL OF THESE FACTORS MAY CREATE STRESS ON THE INSERT MATERIAL AND AFFECT THE DURABILITY OF THE INSERT. BASED ON THE RECEIVED INFO AND THE CONDITION OF THE RETURNED INSERT, IT CANNOT BE DEMONSTRATED THAT THE EVENT WAS CAUSED BY USE. SEE SCANNED PAGES. ADDITIONAL INFO FROM THE IMPORTER REPORT: (B)(4): OT4 INSERT TIP - ACCESSORY TO PIEZOSURGERY BONE CUTTING DEVICE; (B)(6).

Description of Event or Problem · 1

(B)(4): "DURING A PILOT OSTEOTOMY AXIS SURGERY PREPARATION, A PIEZOSURGERY OT4 INSERT TIP FRACTURED, LEAVING THE TIP WEDGED INTO THE BASE OF THE OSTEOTOMY NEAR THE SINUS FLOOR. THE TIP HAD TO BE SURGICALLY REMOVED AT THE IMPLANT SITE FROM A BUCCAL APPROACH".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PIEZOSURGERY OT4 - INSERT TO PIEZOSURGERY ELC MECTRON S.P.A. OT4 1208

Patients

Seq Age Sex Outcome Treatment
1 37 YR Other| R