FDA Adverse Event Injury Summary report: N

VITALITY

MDR report key: 1871192 · Received October 15, 2010

Report

Report Number
2124215-2010-17985
Event Type
Injury
Date Received
October 15, 2010
Date of Event
August 25, 2010
Report Date
October 14, 2010
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
PMA / PMN Number
P960040
Removal / Correction Number
Z-1192-06 THRU Z-1194-06
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS OF TODAY, NO ADDITIONAL INFORMATION IS AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED A BATTERY STATUS OF END OF LIFE (EOL) DUE TO AN EXTENDED CHARGE TIME. THE DEVICE WAS EXPLANTED AND A REQUEST FOR THE RETURN OF THE DEVICE WAS MADE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T180

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention 0158| 4470| 1861| T180| 1853| 4086| 0155