ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2010-17792
- Event Type
- Malfunction
- Date Received
- October 15, 2010
- Date of Event
- August 24, 2010
- Report Date
- August 24, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
TO DATE, INFORMATION SUGGESTS THAT BOTH MEDICAL DEVICES REMAIN ACTIVELY IN SERVICE WITHOUT FURTHER ISSUE. IF NEW INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE FURTHER EVALUATED AND UPDATED.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) IN ASSOCIATION WITH THIS TRANSVENOUS RIGHT VENTRICULAR (RV) RATE/SENSE LEAD PORTION OF THE DEVICE SYSTEM, DID OVERSENSE NON-SUSTAINED VENTRICULAR TACHYCARDIA (NSVT) ACTIVITY WHICH RESULTED IN 2.2 SECONDS OF PAUSE IN PACING FOR THE PACEMAKER DEPENDANT PATIENT. THE PATIENT WAS ASYMPTOMATIC. THE NOISE DID NOT APPEAR TO BE EXTERNAL TO THE SYSTEM AND WAS NOT REPRODUCIBLE. ELECTRICAL RE-PROGRAMMING WAS DONE TO MITIGATE THE ISSUE, AND THE FOLLOWING PHYSICIAN PLANNED TO FOLLOW UP WITH THE PATIENT IN THE NEAR FUTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0185 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | H177| 0185| 4549 |