FDA Adverse Event
Injury
Summary report: N
FINELINE II
MDR report key: 1871153
·
Received October 15, 2010
Report
- Report Number
- 2124215-2010-17601
- Event Type
- Injury
- Date Received
- October 15, 2010
- Date of Event
- August 23, 2010
- Report Date
- August 23, 2010
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE LEAD WAS SURGICALLY ABANDONED AND WILL NOT BE RETURNED FOR TESTING. THEREFORE, BOSTON SCIENTIFIC CRM CANNOT CONFIRM THE REPORTED CLINICAL OBSERVATIONS. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR LEAD WAS EXHIBITING NO CAPTURE AND IMPEDANCE MEASUREMENTS GREATER THAN 2500 OHMS. THEREFORE, A LEAD REVISION WAS PERFORMED AND THE LEAD WAS REMOVED FROM SERVICE AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4470 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | S603| 4469| 4470| 4136| 1290 |