PRIMARY PLUM SET, 15 MICRON FILTER IN SIGHT CHAMBER, CLAVE SECONDARY PORT, 1.2 M
Report
- Report Number
- 9615050-2024-00137
- Event Type
- Malfunction
- Date Received
- February 15, 2024
- Date of Event
- August 7, 2023
- Report Date
- March 21, 2024
- Manufacturer
- ICU MEDICAL COSTA RICA LTD.
- Product Code
- FPA
- PMA / PMN Number
- K141789
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CI
- Reporter Occupation
- 501
Narratives
THE DEVICE IS NOT AVAILABLE FOR EVALUATION; HOWEVER, PHOTOS WERE PROVIDED. INVESTIGATION IS PENDING. E1-INITIAL REPORTER FACILITY NAME: (B)(6) CLINICA.
A VIDEO WAS RETURNED SHOWING A MICRON FILTER LEAKING FROM THE OUTLET FILTER. THE COMPLAINT FILTER LEAKAGE CAN BE CONFIRMED BASED ON THE VIDEO RETURNED BY THE CUSTOMER. ADDITIONALLY, A PICTURE OF THE PACKAGING SHOWING THE LOT NUMBER WAS RETURNED. WITHOUT THE RETURN OF THE USED SAMPLE A COMPREHENSIVE FAILURE INVESTIGATION CANNOT BE PERFORMED, AND A CAUSE CANNOT BE DETERMINED. THE LOT HISTORY WAS REVIEWED, NO NONCONFORMITIES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.
THE EVENT INVOLVED A PRIMARY PLUM SET, LN 15443-31, LOTN 5983801. THE REPORTER STATED THAT THE PHOTOSENSITIVE SET LEAKED IN FILTER AREA WHILE IN USE TO ADMINISTER PARENTERAL NUTRITION TO A PATIENT. THE STATUS OF THE PRODUCT AT THE TIME OF EVENT WAS DURING INFUSION. THERE WERE TWO SETS INVOLVED IN THE EVENT, AND THAT THE SET WAS REMOVED, AND THE ADMINISTRATION OF PARENTAL NUTRITION WAS SUSPENDED. THERE WAS PATIENT INVOLVEMENT, BUT NO REPORT OF PATIENT HARM. THIS CAPTURES ONE OF TWO OCCURRENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2086712 | PRIMARY PLUM SET, 15 MICRON FILTER IN SIGHT CHAMBER, CLAVE SECONDARY PORT, 1.2 M | SET, ADMINISTRATION, INTRAVASCULAR | FPA | ICU MEDICAL COSTA RICA LTD. | 5983801 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | TOTAL PARENTERAL NUTRITION, UNK MFR |