SINGLE USE MECHANICAL LITHOTRIPTOR V
Report
- Report Number
- 9614641-2024-00384
- Event Type
- Malfunction
- Date Received
- February 15, 2024
- Date of Event
- January 29, 2024
- Report Date
- March 11, 2024
- Manufacturer
- AOMORI OLYMPUS CO., LTD.
- Product Code
- LQC
- UDI-DI
- 04953170218422
- PMA / PMN Number
- CLASS2-EXMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, THE TORN GUIDEWIRE TIP WAS LIKELY CAUSE BY THE USER INSERTING THE TIP AND GUIDE WIRE INTO THE ENDOSCOPE AT A SHARP ANGLE AND FORCE WAS APPLIED. HOWEVER, SPECIFIC ROOT CAUSE COULD NOT BE DETERMINED. THE ISSUE CAN BE PREVENTED BY FOLLOWING THE (B)(6) INSTRUCTION MANUAL WHICH PROVIDES THE FOLLOWING WARNING: ¿ WHEN USING THE GUIDE WIRE, INSERT THE INSTRUMENT WITH ITS DISTAL TIP IN PARALLEL WITH THE GUIDE WIRE WHILE HOLDING THE DISTAL TIP AS SHOWN IN FIGURE 4.20. BE CAREFUL NOT TO FORCIBLY INSERT THE INSTRUMENT WITH A SHARP ANGLE BETWEEN THE DISTAL TIP AND THE GUIDE WIRE AS SHOWN IN FIGURE 4.21. THIS MAY DAMAGE THE DISTAL TIP. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
THE DEVICE WAS RETURNED AND THE EVALUATION FOUND THE GUIDEWIRE TIP WAS TORN AND THE MORPHOLOGY OF THE TEAR WAS BRITTLE FRACTURE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS REPORT IS RELATED TO MFR: 00382 (2/2).
IT WAS REPORTED, PART OF THE GUIDEWIRE TIP OF THE SINGLE USE MECHANICAL LITHOTRIPTOR, WHERE THE GUIDEWIRE WAS INSERTED, WAS TORN WHICH MADE IT DIFFICULT TO INSERT IT. THE SECOND ONE WAS USED AS IS AND COMPLETED THE PROCEDURE. THE SAME PROBLEM OCCURRED TWO TIMES IN A ROW. THE ISSUE OCCURRED DURING A THERAPEUTIC PROCEDURE AND IT WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2030031 | SINGLE USE MECHANICAL LITHOTRIPTOR V | SINGLE USE MECHANICAL LITHOTRIPTOR | LQC | AOMORI OLYMPUS CO., LTD. | BML-V442QR-30 | 23K | 04953170218422 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |