FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONRA ANGLE 25LPA
MDR report key: 1871107
·
Received October 12, 2010
Report
- Report Number
- 3003637274-2010-00033
- Event Type
- Injury
- Date Received
- October 12, 2010
- Date of Event
- August 1, 2010
- Report Date
- September 13, 2010
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE PT'S TONGUE WAS BURNED BY THE HEAD OF A HANDPIECE. THE PT WAS NOT GIVEN ANY MEDICATION OR OINTMENT, IT HEALED BY ITSELF. DURING PRODUCT EVAL, LOW DEBRIS LEVEL WAS FOUND IN THE HANDPIECE. THE HEAD WAS DENTED AND THE BACK CAP BUTTON STICKS IN/RUBS ON BEARINGS WHICH CAUSED IT TO HEAT UP THE HEAD OF THE HANDPIECE.
Description of Event or Problem · 1
A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE PT WAS BURNED BY THE HEAD OF A HANDPIECE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONRA ANGLE 25LPA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |