FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 1ML LL BNS

MDR report key: 18710427 · Received February 15, 2024

Report

Report Number
1213809-2024-00085
Event Type
Malfunction
Date Received
February 15, 2024
Date of Event
January 26, 2024
Report Date
May 16, 2024
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
50382903096485
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B.3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

TWO PHOTOS OF A 1ML LUER-LOK SYRINGE WERE RECEIVED AND EVALUATED. BOTH IMAGES SHOW A LOOSE SYRINGE FROM TWO DIFFERENT ANGLES, WITH THE STOPPER JAMMED BETWEEN THE BARREL AND THE PLUNGER ROD. A SAMPLE OF A 1 ML LUER-LOK SYRINGE WAS RECEIVED AND EVALUATED. THE SYRINGE WAS RECEIVED DISASSEMBLED. THE BARREL WAS LOOSE, AND THE PLUNGER ROD HAD A DEFORMED STOPPER CONSISTENT WITH A DISTORTED CONDITION. THE POTENTIAL ROOT CAUSE OF THE DISTORTED STOPPER DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED BATCH NUMBER 1200757 SHOWING NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SYRINGE 1ML LL BNS PLUNGER ROD WAS BROKEN/DAMAGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER TRANSLATED FROM DUTCH TO ENGLISH: PLUNGER OF SYRINGE IS DAMAGED PLUNGER DAMAGED.

Description of Event or Problem · 0

PLUNGER OF SYRINGE IS DAMAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
543204 BD SYRINGE 1ML LL BNS PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 1200757 50382903096485

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown