FDA Adverse Event Injury Summary report: N

TALENT STENT GRAFT SYSTEM WITH XCELERANT - HYDRO

MDR report key: 1870994 · Received October 8, 2010

Report

Report Number
2953200-2010-01919
Event Type
Injury
Date Received
October 8, 2010
Date of Event
September 8, 2010
Report Date
September 8, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL, RESULTS: TREATMENT OF PRE-EXISTING RUPTURED ABDOMINAL AORTIC ANEURYSM, THE AORTIC NECK WAS SHORT IN LENGTH (9MM), SEVERELY ANGULATED AND HAD A REVERSE FUNNEL, 21MM PROXIMALLY TO DISTALLY 24MM IN DIAMETER. ENDOLEAK. THE EXPLANTED STENT WAS DISCARDED, UNABLE TO F/U. EVAL, CONCLUSIONS: TREATMENT OF PRE-EXISTING RUPTURED ABDOMINAL AORTIC ANEURYSM, THE AORTIC NECK WAS SHORT IN LENGTH (9MM), SEVERELY ANGULATED AND HAD A REVERSE FUNNEL, 21MM PROXIMALLY TO DISTALLY 24MM IN DIAMETER.

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED EMERGENTLY IN A PT FOR THE ENDOVASCULAR TREATMENT OF A RUPTURED ABDOMINAL AORTIC ANEURYSM 1 MONTH AGO. ANEURYSM SIZE WAS NOT REPORTED. VESSEL MORPHOLOGY WAS REPORTED AS THE AORTIC NECK WAS SHORT IN LENGTH (9MM), SEVERELY ANGULATED AND HAD A REVERSE FUNNEL, 21MM PROXIMALLY TO DISTALLY 24MM IN DIAMETER. THE STENT GRAFTS WERE SUCCESSFULLY IMPLANTED, THE FINAL ANGIOGRAM SHOWED NO ENDOLEAKS. IT WAS REPORTED THAT LATER THAT SAME DAY THE PT WAS BROUGHT BACK TO THE OPERATING ROOM AND THERE WAS A PROXIMAL TYPE I ENDOLEAK. THE PHYSICIAN ELECTED TO CONVERT THE PT TO AN OPEN SURGICAL REPAIR, DUE TO THE SHORT LENGTH AND SEVERE ANGULATION OF THE AORTIC NECK. THE EXPLANTED STENT GRAFTS WERE DISCARDED BY THE USER FACILITY. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT STENT GRAFT SYSTEM WITH XCELERANT - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00426783

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention