ENDOWRIST
Report
- Report Number
- 2955842-2024-11453
- Event Type
- Malfunction
- Date Received
- February 15, 2024
- Date of Event
- January 21, 2024
- Report Date
- January 24, 2024
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- PMA / PMN Number
- K214095
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- 003
Narratives
A RETURN MATERIAL AUTHORIZATION (RMA)) WAS ISSUED TO EVALUATE THE INTUITIVE SURGICAL, INC. (ISI) DEVICE. ADDITIONAL INFORMATION IS BEING GATHERED TO DETERMINE THE CONTRIBUTION OF THE DEVICE TO THE CUSTOMER REPORTED ISSUE.
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE LARGE SUTURECUT NEEDLE DRIVER INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A BROKEN GRIP CABLE AT THE IDLER PULLEYS. THE CABLE WAS FULLY BROKEN. THERE WAS NO EVIDENCE OF DISCOLORATION OR CORROSION/CONTAMINATION ON THE CABLES TO INDICATE A REPROCESSING INDUCED ISSUE. THE COMPONENTS SURROUNDING THE BROKEN CABLE DID NOT EXHIBIT ABNORMAL SHARP EDGES OR DAMAGE THAT WOULD HAVE CONTRIBUTED TO THE CABLE BREAKING. ADDITIONALLY, THE INSTRUMENT WAS FOUND TO HAVE BLADE DAMAGE AT THE MIDPOINT OF THE BLADE. BOTH BLADE EDGES WERE INDENTED. THE CUTTING EDGE DID NOT EXHIBIT ANY SIGNS OF CORROSION THAT WOULD HAVE CONTRIBUTED TO THE BLADE DAMAGE. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED VENTRAL HERNIA TAPP SURGICAL PROCEDURE, THE LARGE SUTURE CUT NEEDLE DRIVER WAS FOUND TO HAVE BROKEN CABLE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2159048 | ENDOWRIST | LARGE SUTURE CUT NEEDLE DRIVER | NAY | INTUITIVE SURGICAL, INC | 471296-07 | T10200729 0015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |