FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONRA ANGLE 25LPA
MDR report key: 1870867
·
Received October 7, 2010
Report
- Report Number
- 3003637274-2010-00030
- Event Type
- Injury
- Date Received
- October 7, 2010
- Date of Event
- September 1, 2010
- Report Date
- September 10, 2010
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
WHILE WORKING ON THE TOOTH# 30, THE PATIENT CHEEK WAS BURNED. BURN AREA WAS THE SIZE OF DIME. THE PATIENT WAS NOT GIVEN ANY MEDICATION OR OINTMENT, IT HEALED BY ITSELF. DURING PRODUCT EVALUATION, MEDIUM DEBRIS LEVEL WAS FOUND IN THE HANDPIECE. THE BEARINGS WHERE GRITTY AND THE HEAD WAS DENTED AT 11. THE DENT KEEPS THE BACK CAP DEPRESSED WHICH LEAD TO HEATS UP THE HEAD.
Description of Event or Problem · 1
A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PATIENT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE PATIENT CHEEK WAS BURNED BY THE HEAD OF A HANDPIECE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONRA ANGLE 25LPA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |