FDA Adverse Event Injury Summary report: N

GENTLEPOWER LUX CONRA ANGLE 25LPA

MDR report key: 1870867 · Received October 7, 2010

Report

Report Number
3003637274-2010-00030
Event Type
Injury
Date Received
October 7, 2010
Date of Event
September 1, 2010
Report Date
September 10, 2010
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WHILE WORKING ON THE TOOTH# 30, THE PATIENT CHEEK WAS BURNED. BURN AREA WAS THE SIZE OF DIME. THE PATIENT WAS NOT GIVEN ANY MEDICATION OR OINTMENT, IT HEALED BY ITSELF. DURING PRODUCT EVALUATION, MEDIUM DEBRIS LEVEL WAS FOUND IN THE HANDPIECE. THE BEARINGS WHERE GRITTY AND THE HEAD WAS DENTED AT 11. THE DENT KEEPS THE BACK CAP DEPRESSED WHICH LEAD TO HEATS UP THE HEAD.

Description of Event or Problem · 1

A DENTIST WAS PERFORMING A ROUTINE PROCEDURE ON A PATIENT USING A DENTAL ELECTRICAL HANDPIECE WHEN THE PATIENT CHEEK WAS BURNED BY THE HEAD OF A HANDPIECE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX CONRA ANGLE 25LPA DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention