STRATA II, ADJUSTABLE DELTA VALVE, SMALL
Report
- Report Number
- 2021898-2010-00244
- Event Type
- Injury
- Date Received
- October 7, 2010
- Date of Event
- September 1, 2010
- Report Date
- September 13, 2010
- Manufacturer
- MEDTRONIC NEUROSURGERY
- Product Code
- JXG
- PMA / PMN Number
- K042465
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SE
- Reporter Occupation
- NURSE
Narratives
(B)(4). THE RETURNED VALVE WAS PATENT, PASSED LEAK AND REFLUX TESTING AND MET ALL SPECS FOR PRESSURE-FLOW AND PREIMPLANTATION TESTING AT 0 CM HYDROSTATIC PRESSURE. HOWEVER, IT DID NOT MEET THE REQUIREMENTS FOR SIPHON TESTING, WHICH PRECLUDED MEASURING OF PRESSURE-FLOW AT -50 CM HYDROSTATIC PRESSURE. PROTEINACEOUS DEBRIS WAS FOUND ON THE INTERIOR OF THE VALVE, PARTICULARLY INSIDE THE DELTA CHAMBER. DEBRIS WITHIN THE VALVE MAY INTERFERE WITH THE ANTI-SIPHON DEVICE, RESULTING IN SIPHONING. ALL PERFORMANCE LEVELS COULD BE SET WITH A SINGLE ATTEMPT, AND A LEVEL CHANGE COULD NOT BE INDUCED BY LABORATORY PERSONNEL WITHOUT USING A MAGNET. A DISSECTION OF THE ADJUSTMENT MECHANISM SHOWED NO ANOMALIES. STRATA PRODUCTS ARE DESIGNED TO BE ADJUSTED BY A STRONG MAGNET. INADVERTENT ADJUSTMENTS BY EXTERNAL MAGNETS ARE A KNOWN COMPLICATION WITH ADJUSTABLE VALVES. THE INSTRUCTIONS FOR USE THAT ACCOMPANIES THE PRODUCT CAUTIONS THAT DEVICES KNOWN TO CONTAIN MAGNETS SHOULD BE KEPT AWAY FROM THE IMMEDIATE VALVE IMPLANT LOCATION, AS THEY MAY HAVE AN EFFECT ON THE PERFORMANCE LEVEL SETTING OF THE STRATA-TYPE VALVE. ALL MAGNETS HAVE AN EXPONENTIALLY DECREASING EFFECT ON THE VALVE THE FURTHER AWAY THEY ARE LOCATED. WHILE IT IS IMPOSSIBLE TO QUANTIFY ALL THE VARIOUS MAGNETIC FIELDS IN OUR ENVIRONMENT, THE VAST MAJORITY WILL NOT IMPACT THE VALVE AS LONG AS DISTANCE FROM THE VALVE IMPLANT IS MAINTAINED. A REVIEW OF THE MFG RECORDS SHOWED NO ANOMALIES. ALL OF OUR VALVES ARE 100% TESTED AT THE TIME OF MANUFACTURE.
THE PRESSURE LEVEL CHANGED FROM 1.5 TO 1.0 AND THEY CLAIM THAT THIS CHANGE IN SETTING HAPPENED WHEN THIS PT ENTERED OR LEFT A STORE WALKING THROUGH THE METAL DETECTOR. THE DEVICE WAS EXPLANTED AND REPLACED WITH A FIXED PRESSURE VALVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STRATA II, ADJUSTABLE DELTA VALVE, SMALL | JXG | MEDTRONIC NEUROSURGERY | NA | 46751 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization| R |