VENTAK PRIZM
Report
- Report Number
- 2124215-2010-17106
- Event Type
- Malfunction
- Date Received
- October 15, 2010
- Date of Event
- August 25, 2010
- Report Date
- August 25, 2010
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ERI AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE ERI CHARGE TIME LIMIT OF 18 SECONDS. THE DEVICE PASSED A SERIES OF DIAGNOSTIC TESTS TO ASSESS THE PERFORMANCE OF DEFIBRILLATION, PACING, SENSING, HIGH VOLTAGE SHOCKING, AND DIAGNOSTIC RECORDING FUNCTIONS. THE DEVICE'S MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE AND PROGRAMMED SETTINGS. CHARGE TIMES IN EXCESS OF THE 18 SECOND ERI CHARGE TIME LIMIT TRIGGERED ERI EARLIER THAN EXPECTED. THIS WAS DUE TO A BUILD-UP OF INTERNAL BATTERY IMPEDANCE, PREVENTING THE DEVICE FROM MEETING ITS EXPECTED LONGEVITY PER DEVICE LABELING. THIS ISSUE IS DISCUSSED IN OUR Q3 2010 CRM PRODUCT PERFORMANCE REPORT.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE WAS ELECTIVELY EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THERE WERE NO ALLEGATIONS OR COMPLAINTS AGAINST THE DEVICE'S OPERATION, FUNCTIONALITY OR LONGEVITY. UPON RETURN, INITIAL ROUTINE ANALYSIS REVEALED THE DEVICE HAD FAILED TO MEET LONGEVITY SPECIFICATIONS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VENTAK PRIZM | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | 1852 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | F102| 1852 |