PERCUTANEOUS TRIAL LEAD
Report
- Report Number
- 1627487-2010-02795
- Event Type
- Injury
- Date Received
- October 6, 2010
- Date of Event
- September 10, 2010
- Report Date
- September 10, 2010
- Manufacturer
- ADVANCED NEUROMODULATION SYSTEMS, INC.
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVALUATION: METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE ALSO REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED AND WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. ANS HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. ANS DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
ON (B)(6) 2010, THE PT WAS IMPLANTED WITH A TRIAL SCS LEAD. ON (B)(6) 2010, IT WAS REPORTED THAT THE PT SNEEZED AND THE ANCHOR RIPPED OFF HIS SKIN. THE PT'S LEAD SUBSEQUENTLY MIGRATED AND WAS EXPLANTED ON (B)(6) 2010.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCUTANEOUS TRIAL LEAD | SPINAL CORD STIMULATION LEAD | LGW | ADVANCED NEUROMODULATION SYSTEMS, INC. | 3086 | 3170799 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |