FDA Adverse Event
Injury
Summary report: N
HAKIM VALVE
MDR report key: 1870818
·
Received October 6, 2010
Report
- Report Number
- 1226348-2010-00321
- Event Type
- Injury
- Date Received
- October 6, 2010
- Manufacturer
- CODMAN & SHURTLEFF, INC. MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- K974739
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CODMAN HAS REQUESTED RETURN OF THE VALVE FOR EXAMINATION. HISTORICALLY, FOR COMPLAINTS FOR THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE ACCUMULATIONS OF BIOLOGICAL DEBRIS WITHIN THE VALVES, WHICH HAVE RESULTED IN BLOCKAGES WITHIN THE DEVICES. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE, OR, IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CLOSED AT THIS TIME.
Description of Event or Problem · 1
AFFILIATE REPORTED THE FLUID DID NOT FLOW THROUGH THE DEVICE AND THE DEVICE WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HAKIM VALVE | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC. MEDOS S.A. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |