RESTORE RECHARGEABLE NEUROSTIMULATOR
Report
- Report Number
- 3004209178-2010-07881
- Event Type
- Injury
- Date Received
- October 12, 2010
- Date of Event
- April 6, 2006
- Report Date
- February 13, 2009
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO., JUNCOS
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). REASON FOR LATE MDR DUE TO IMPLEMENTATION OF PROCESS IMPROVEMENT.
FOLLOWING IMPLANT, IT WAS REPORTED THAT THE PT NO LONGER HAD ADEQUATE PARESTHESIA COVERAGE AND THE IMPEDANCE OF ONE HALF OF THE ELECTRODES WAS ABNORMALLY HIGH. THE DEVICE WAS REPROGRAMMED "MANY" TIMES OVER A PERIOD OF "SEVERAL" DAYS, BUT THIS DID NOT RESOLVE THE PROBLEM. AN X-RAY WAS TAKEN REVEALING THERE WAS NO DISCONTINUITY OF THE SYSTEM, BUT THE EPIDURAL ELECTRODE WAS LOWER THAN PREVIOUSLY. THIS WAS NOT UNEXPECTED SINCE IT WAS INTENTIONALLY PLACED LOWER. BEFORE A REVISION COULD BE PERFORMED, THE PT'S PAIN BECAME INTOLERABLE, AND THE PT WENT TO THE EMERGENCY ROOM FOR EMERGENCY PAIN MGMT. THE PT WAS ADMITTED FOR PARENTERAL MEDICATION THERAPY OF HIS REFRACTORY PAIN CONDITION . THE DEVICE WAS REVISED AND THE PT EXPERIENCED PAIN RELIEF AND PARESTHESIA COVERAGE THAT WAS "BETTER THAN EVER". FOR INFO ABOUT EVENTS THAT OCCURRED BEFORE THE IMPLANT, SEE MFR REPORT # 3004209178-2010-07879.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESTORE RECHARGEABLE NEUROSTIMULATOR | LGW | MDT PUERTO RICO OPERATIONS CO., JUNCOS | 37711 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Hospitalization| R | EXTENSION: MODEL 37083, LOT# NKC006621N| IMPLANTED:| EXPLANTED:| IMPLANTED:| LEAD: MODEL 3998, LOT# J0437134V| EXPLANTED:| EXPLANTED:| IMPLANTED:| PROGRAMMER: MODEL 37742, LOT# NJD022094N| EXPLANTED:| EXTENSION: MODEL 37083, LOT# NKC006620N| EXPLANTED:| ACCESSORY: MODEL 37752, LOT# NKA017140N| IMPLANTED:| IMPLANTED: |